Maximum Permitted Error in Thermocouples - °C  (IEC 584-2, 1982)
Temperature B type E type J type K type N type R type S type T type
-200 - - - 3.0 3.0 - - 3.0
-100 - - - 2.5 2.5 - - 1.5
0 - 1.7 1.5 1.5 1.5 1.0 1.0 0.5
200 - 1.7 1.5 1.5 1.5 1.0 1.0 0.8
400 - 2.0 1.6 1.6 1.6 1.0 1.0 -
600 1.5 3.0 2.4 2.4 2.4 1.0 1.0 -
800 2.0 4.0 - 3.2 3.2 1.0 1.0 -
1000 2.5 - - 4.0 4.0 1.0 1.0 -
1200 3.0 - - 9.0 9.0 1.3 1.3 -
1400 3.5 - - - - 1.9 1.9 -
1600 4.0 - - - - 2.5 2.5 -

Copyright © 2008 Control Automation, Inc. All rights reserved.